Connect Systems CS1000 Specifications Page 60

  • Download
  • Add to my manuals
  • Print
  • Page
    / 112
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 59
ContaminationSensor CS 1000 Reading the analog output
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 60/112
BeWa CS1000 3764916 300 en-us 2012-08-29.doc 2012-08-29
HDA.SAE – Analog signal SAE to the HDA 5500
The HDA.SAE signal consists of 6 values (START / SAE A / SAE B / SAE C / SAE
D / Status) which are output sequentially. Synchronization with the downstream
control unit is a prerequisite.
The signal output is as follows:
Time Measured
variable
Signal duration
in s
Current/Voltage
Start signal 0 -- 2 20 mA / 10 V
Pause 2 4 mA / 2 V
Signal 1 SAE A 2 Current / Voltage for signal
Pause 2 4 mA / 2 V
Signal 2 SAE B 2 Current / Voltage for signal
Pause 2 4 mA / 2 V
Signal 3 SAE C 2 Current / Voltage for signal
Pause 2 4 mA / 2 V
Signal 4 SAE D 2 Current / Voltage for signal
Pause 2 4 mA / 2 V
Signal 5 Status 2 Current / Voltage for signal
Pause 30 4 mA / 2 V
Page view 59
1 2 ... 55 56 57 58 59 60 61 62 63 64 65 ... 111 112

Comments to this Manuals

No comments